Table 1 Monitoring procedures during abdominal normothermic oxygenated recirculation

From: Development of a preclinical model of donation after circulatory determination of death for translational application

Monitored parameters Objectives Type of monitoring Corrective actions
Pump flow rate 2.5 to 3 l/min Continuous monitoring Vascular filling: saline and hydroxyethyl starch (10 ml/(kg h))
Mean arterial pressure >60 mmHg Continuous infusion of noradrenalin on the circuit
Central venous pressure <5 mmHg
pH 7.5 Blood samples on the circuit every 30 minutes Continuous infusion of sodium bicarbonate on the circuit
Partial pressure of carbon dioxide 30 to 40 mmHg Sweeping gas flow adjustment
Partial pressure of oxygen 100 to 150 mmHg Fraction of inspired oxygen adjustment
Activated clotting time >200 s   Bolus of 5000 IU unfractionated heparin on the circuit